Grayscale Values in Cone Beam Computed Tomography: Scope and Limitations

Authors

  • Ana Cecilia Ruiz-Imbert Especialista en Radiología Oral y Maxilofacial, Docente del Departamento de Ciencias Diagnósticas y Quirúrgicas, Facultad de Odontología, Universidad de Costa Rica, San José, Costa Rica. Author https://orcid.org/0000-0002-0700-6814
  • Deivi Cascante-Sequeira Estudiante de la Maestría en Radiología Oral, Escuela Dental de Piracicaba, Universidad de Campinas, São Paulo, Brasil Author https://orcid.org/0000-0002-5009-6632

DOI:

https://doi.org/10.15517/ijds.2021.45106

Keywords:

Cone-beam volumetric imaging; Cone-beam computed tomography or cone-beam volumetric tomography; Hounsfield units; grey levels.

Abstract

Cone-beam computed tomography (CBCT) is a 3D imaging technique widely used in maxillofacial diagnosis. The grayscale value (GSV) is a number that represents the amount of attenuation of the X-ray beam by the material contained in each voxel or structural unit of the tomographic volume. Similarly, in computed tomography (CT) used in medical radiology, the attenuation values are standardized in the Hounsfield Unit (HU) scale. Although GSV may have interesting potential applications in maxillofacial diagnosis, it is essential to know that HU differ from GSV. The latter are susceptible to multiple technical factors during the tomographic acquisition, so their value can vary among different CBCT scanners or when technical parameters are modified. Hence, GSV should not be extrapolated between different CBCT machines, and their use should be cautious while more investigation is available considering various equipment and acquisition protocols.

Downloads

Download data is not yet available.

References

Molteni R. The way we were (and how we got here): fifty years of technology changes in dental and maxillofacial radiology.

Dentomaxillofacial Radiol. 2020; 20200133.

Pauwels R., Jacobs R., Singer S.R., Mupparapu M. CBCT-based bone quality assessment: Are Hounsfield units applicable? Dentomaxillofacial Radiol. 2015; 44 (1). DOI: https://doi.org/10.1259/dmfr.20140238

Gaêta-Araujo H., Alzoubi T., Vasconcelos K. de F., Orhan K., Pauwels R., Casselman J.W., et al. Cone-beam computed tomography in dentomaxillofacial radiology: a two-decade overview. Dentomaxillofacial Radiol. 2020; 49 (8): 20200145. DOI: https://doi.org/10.1259/dmfr.20200145

Azeredo F., De Menezes L.M., Enciso R., Weissheimer A., De Oliveira R.B. Computed gray levels in multislice and cone-beam computed tomography. Am J Orthod Dentofac Orthop. 2013;144 (1): 147-55. DOI: https://doi.org/10.1016/j.ajodo.2013.03.013

Pauwels R., Nackaerts O., Bellaiche N., Stamatakis H., Tsiklakis K., Walker A., et al. Variability of dental cone beam CT grey values for density estimations. Br J Radiol. 2013; 86 (1021): 1-9. DOI: https://doi.org/10.1259/bjr.20120135

Miles D.A., Danforth R.A. A Clinician’s Guide to Understanding Cone Beam Volumetric Imaging (CBVI) Educational Objectives. Available from: www.ada.org/goto/cerp.%5Cnwww.ineedce.com

Pauwels R., Araki K., Siewerdsen J.H., Thongvigitmanee S.S. Technical aspects of dental CBCT: State of the art. Vol. 44, Dentomaxillofacial Radiology. British Institute of Radiology; 2015. DOI: https://doi.org/10.1259/dmfr.20140224

Goulston R., Davies J., Horner K., Murphy F. Dose optimization by altering the operating potential and tube current exposure time product in dental cone beam CT: A systematic review. Dentomaxillofacial Radiol. 2016; 45 (3). DOI: https://doi.org/10.1259/dmfr.20150254

Oenning A.C., Jacobs R., Pauwels R., Stratis A., Hedesiu M., Salmon B. Cone-beam CT in paediatric dentistry: DIMITRA project position statement. Vol. 48, Pediatric Radiology. Springer Verlag; 2018. p. 308-16. DOI: https://doi.org/10.1007/s00247-017-4012-9

Mutalik S., Tadinada A., Molina M.R., Sinisterra A., Lurie A. Effective doses of dental cone beam computed tomography: effect of 360-degree versus 180-degree rotation angles. Oral Surg Oral Med Oral Pathol Oral Radiol. 2020; 130 (4): 433-46. DOI: https://doi.org/10.1016/j.oooo.2020.04.008

Sedentexct. Radiation Protection 172: Cone Beam CT for Dental and Maxillofacial Radiology-Evidence-based Guidelines. Off Off Publ Eur Communities [Internet]. 2012; 156. Available from: www.sedentexct.eu

Schulze R., Heil U., Groß D., Bruellmann D.D., Dranischnikow E., Schwanecke U., et al. Artefacts in CBCT: A review. Dentomaxillofacial Radiol. 2011; 40 (5): 265-73. DOI: https://doi.org/10.1259/dmfr/30642039

Nagarajappa A., Dwivedi N., Tiwari R. Artifacts: The downturn of CBCT image. J Int Soc Prev Community Dent. 2015; 5 (6): 440. DOI: https://doi.org/10.4103/2231-0762.170523

Candemil A.P., Salmon B., Freitas D.Q., Ambrosano G.M.B., Haiter-Neto F., Oliveira M.L. Metallic materials in the exomass impair cone beam CT voxel values. Dentomaxillofacial Radiol. 2018; 47 (6): 2-4. DOI: https://doi.org/10.1259/dmfr.20180011

Molteni R. Prospects and challenges of rendering tissue density in Hounsfield units for cone beam computed tomography. Oral Surg Oral Med Oral Pathol Oral Radiol [Internet]. 2013; 116 (1): 105-19. DOI: https://doi.org/10.1016/j.oooo.2013.04.013

Parsa A., Ibrahim N., Hassan B., Motroni A., Der Van Stelt P., Wismeijer D. Influence of cone beam CT scanning parameters on grey value measurements at an implant site. Dentomaxillofacial Radiol. 2013; 42 (3). DOI: https://doi.org/10.1259/dmfr/79884780

Bujtár P., Simonovics J., Zombori G., Fejer Z., Szucs A., Bojtos A., et al. Internal or in-scan validation: A method to assess CBCT and MSCT gray scales using a human cadaver. Oral Surg Oral Med Oral Pathol Oral Radiol [Internet]. 2014; 117 (6): 768-79. DOI: https://doi.org/10.1016/j.oooo.2014.02.012

Reeves T.E., Mah P., McDavid W.D. Deriving Hounsfield units using grey levels in cone beam CT: A clinical application. Dentomaxillofacial Radiol. 2012; 41 (6): 500-8. DOI: https://doi.org/10.1259/dmfr/31640433

Mah P., Reeves T.E., McDavid W.D. Deriving Hounsfield units using grey levels in cone beam computed tomography. Dentomaxillofacial Radiol. 2010; 39 (6): 323-35. DOI: https://doi.org/10.1259/dmfr/19603304

Razi T., Niknami M., Alavi Ghazani F. Relationship between Hounsfield Unit in CT Scan and Gray Scale in CBCT. J Dent Res Dent Clin Dent Prospects [Internet]. 2014; 8 (2): 107-10.

Nomura Y., Watanabe H., Shirotsu K., Honda E., Sumi Y., Kurabayshi T. Stability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral content. Clin Oral Implants Res. 2013; 24 (5): 543-8. DOI: https://doi.org/10.1111/j.1600-0501.2012.02420.x

Guerra E.N.S., Almeida F.T., Bezerra F.V., Figueiredo P.T.D.S., Silva M.A.G., De Luca Canto G., et al. Capability of CBCT to identify patients with low bone mineral density: A systematic review. Dentomaxillofacial Radiol. 2017; 46 (8). DOI: https://doi.org/10.1259/dmfr.20160475

Çolak M. An evaluation of bone mineral density using cone beam computed tomography in patients with ectodermal dysplasia: A retrospective study at a single center in Turkey. Med Sci Monit. 2019; 25: 3503-9. DOI: https://doi.org/10.12659/MSM.914405

Hakim S.G., Glanz J., Ofer M., Steller D., Sieg P. Correlation of cone beam CT-derived bone density parameters with primary implant stability assessed by peak insertion torque and periotest in the maxilla. J Cranio-Maxillofacial Surg [Internet]. 2019; 47 (3): 461-7. DOI: https://doi.org/10.1016/j.jcms.2019.01.002

Kaya S., Yavuz I., Uysal I., Akku Z. Measuring bone density in healing periapical lesions by using cone beam computed tomography: A clinical investigation. J Endod. 2012; 38 (1): 28-31. DOI: https://doi.org/10.1016/j.joen.2011.09.032

Komori M., Miuchi S., Hyodo J., Kobayashi T., Hyodo M. The gray scale value of ear tissues undergoing volume-rendering high-resolution cone-beam computed tomography. Auris Nasus Larynx [Internet]. 2018; 45 (5): 971-9. DOI: https://doi.org/10.1016/j.anl.2018.01.012

Emadi N., Safi Y., Akbarzadeh Bagheban A., Asgary S. Comparison of CT-number and gray scale value of different dental materials and hard tissues in CT and CBCT. Iran Endod J. 2014; 9 (4): 283-6.

Oliveira M.L., Tosoni G.M., Lindsey D.H., Mendoza K., Tetradis S., Mallya S.M. Assessment of CT numbers in limited and medium field-of-view scans taken using accuitomo 170 and veraviewepocs 3De cone-beam computed tomography scanners. Imaging Sci Dent. 2014; 44 (4): 279-85. DOI: https://doi.org/10.5624/isd.2014.44.4.279

Shokri A., Ramezani L., Bidgoli M. Akbarzadeh M., Ghazikhanlu-Sani K., Fallahi-Sichani H. Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans.

Imaging Sci Dent. 2018; 48 (1): 31-9. DOI: https://doi.org/10.5624/isd.2018.48.1.31

Martins L.A.C., Queiroz P.M., Nejaim Y., De Faria Vasconcelos K., Groppo F.C., Haiter-Neto F. Evaluation of metal artefacts for two CBCT devices with a new dental arch phantom. Dentomaxillofacial Radiol. 2020; 49 (5). DOI: https://doi.org/10.1259/dmfr.20190385

Machado A.H., Fardim K.A.C., De Souza C.F., Sotto-Maior B.S., Assis N.M.S.P., Devito K.L. Effect of anatomical region on the formation of metal artefacts produced by dental implants in cone beam computed tomographic images. Dentomaxillofacial Radiol. 2018; 47 (3). DOI: https://doi.org/10.1259/dmfr.20170281

Candemil A.P., Salmon B., Freitas D.Q., Haiter-Neto F., Oliveira M.L. Distribution of metal artifacts arising from the exomass in small field-of-view cone beam computed tomography scans. Oral Surg Oral Med Oral Pathol Oral Radiol [Internet]. 2020; 130 (1): 116-25. DOI: https://doi.org/10.1016/j.oooo.2020.01.002

Magill D., Beckmann N., Felice M.A., Yoo T., Luo M., Mupparapu M. Investigation of dental cone-beam CT pixel data and a modified method for conversion to hounsfield unit (HU). Dentomaxillofacial Radiol. 2018; 47 (2). DOI: https://doi.org/10.1259/dmfr.20170321

Published

2026-06-15